T2030/T2037: 6 µl; typical elution volumes are 6-20 µl
T2040/T2047: 20 µl; typical elution volumes are 20-50 µl
T2050/T2057: 50 µl; typical elution volumes are 50-100 µl
Using lower elution volumes will likely result in variance due to inconsistent wetting of the matrix. Yield may slightly increase if a larger volume is used, but the RNA will be less concentrated.
*A second elution can be performed to maximize RNA recovery. This second elution can be performed using a fresh aliquot of nuclease-free water or the eluent from the first elution to maximize recovery while minimizing volume.