Troubleshooting Guide for use with SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit (NEB #E2019)
Troubleshooting Guide for use with SARS-CoV-2 Rapid Colorimetric LAMP Assay Kit (NEB #E2019)
Note: For additional assistance please refer to the product FAQ
PROBLEM | POSSIBLE CAUSE(S) | SOLUTION(S) |
Colorimetric LAMP reaction is orange prior to amplification | Nucleic acid sample input is incompatible |
|
Repeated exposure of the colorimetric LAMP master mix to atmosphere causes a slight pH drop |
|
|
Positive Control reaction fails to turn yellow following incubation at 65°C | Improper pipetting during colorimetric LAMP set-up |
|
Poor mixing of reagents during set-up |
|
|
NTC reaction turns yellow following incubation at 65°C | Reagents are contaminated with the positive control |
|
Improper pipetting during colorimetric LAMP set-up |
|
|
Internal Control reaction fails to turn yellow following incubation at 65°C | Improper pipetting during colorimetric LAMP set-up |
|
Sample nucleic acid is incompatible with colorimetric LAMP |
|
|
Sample nucleic acid is isolated from a non-human source |
|
|
Poor mixing of reagents during set-up |
|
|
RNA template is contaminated or degraded |
|